Scopy
Table of Contents
1. Time-Resolved Spetroscopy
First proposed by Ahmed H. Zewail
Mechanical shutter was used back in 1960. They work up to few milisecond to microsecond. Electronic control caps at Gigahertz. Now use pulse-lazer upto attosecond resolution.
1.1. Pump-probe method
The pump–probe method is a basic technique in time-resolved spectroscopy for watching very fast processes, often on femtosecond to picosecond timescales.
- A short, intense pump pulse hits the sample first.
- It excites the system, for example promoting an electron to a higher energy state.
- After a controlled delay, a second weaker probe pulse arrives.
- The probe checks how the system has changed after that delay.
- By repeating the experiment with different time delays, build up a picture of the dynamics in time.
Pump–probe spectroscopy can reveal:
- Excited-state lifetimes
- Relaxation and energy transfer
- Vibrational motion
- Chemical reaction dynamics
- Electron transfer
Depending on the method, the probe may:
- measure absorption changes,
- induce fluorescence,
- cause stimulated emission,
- or produce ionization.
1.1.1. Laser-Induced Fluorescence (LIF)
The pump excites the molecule, and the detected fluorescence gives information about the excited state and its decay.
1.1.2. Resonance two-photon ionization (R2PI)
One photon excites the molecule to an intermediate state, and another photon ionizes it. This is useful for sensitive detection of specific species and states.
Use pulsed nozzle to spary the sample, and the timeing is matched with the YAG laser-dye laser
Monitored using
- Etalon: Fabry-Perot Cavity
- I₂ LIF
- Wavelength monitor
2. Absorption Spectroscopy
2.1. Cavity Ring-down Spectroscopy
- CRDS
- Cavity Ring-down Laser Absorption Spectroscopy (CRLAS)
Laser is resonated within a cavity causing absorption by the sample (normally, in gas phase) within the cavity.
High-finesse optical catity is used, which effectively create tens of kilometers of optical path length. Due to the long path length, this is used in detection of miniscule concentration.
The signal is produced after the laser is turned off, and it decays exponentially \[ I(t) = I_0 e^{\frac{t}{\tau(\lambda)}}. \] Broadband laser is used for broader spectral coverage. The mean lifetime is used to measure the concentration.
American company Piccaro Technology is manufacturing this commercially.
3. Raman Spectroscopy
- IR은 공간 분해능이 높지 않지만 Raman은 높다 (300 nm).
- Raman은 간접적인 빛을 측정하기 때문에 직접적인 빛을 측정하는 IR에 비해 해석이 어렵다. 따라서 고분자에서는 IR이 더 유용해진다.
- 라만은 경험이 있어야 다룰 수 있다.
Sotkes Raman scattering First excites into virtual energy state (not stable state) and falls back to a higher vibrational state. Stokes shift occurs that elongates the wavelenth
The virtual state here means the transitional non-stable state obtained from perturbation theory as a superposition of energy eigenstates, although the true state would be an eletron-photon state.
Anti-Stokes Raman scattering First excites into virtual energy state and falls back to a lower vibrational state. Stokes shift occurs that shorten the wavelenth
The strength of Raman signal is extremely weak (1 in 10 million) compared to Rayleigh scattering.
Manufacturers
- IDEX Health and Science
- Semrock (optical fibers)
- Horiba (Raman Spectroscopy)
- Park System
3.1. Raman Efficiency
\[ I(\omega_s) = \frac{\omega_s^4}{12\pi\varepsilon_0c^3}Q_0^2|A|^2 \left| \frac{\delta \alpha}{\delta Q} \right|^2 \]
3.2. Depth of Penetration
High frequency penetrates less.
3.3. Surface Enhanced Raman Spectroscopy
- SERS
Molecules are adsorbed on rough metal surfaces or by nanostructures such as plasmonic-magnetic silica nanotubes.
The signal is amplified 10⁶ to 10¹² times on those special surface.
Local surface plasmon resonance (LSPR) create a dense electric field region, at which the signal is amplified by the factor of 10⁶ to 10⁸. \[ \text{signal} \propto \left| \frac{E_{\rm local}}{E_0} \right|^4. \]
Universal SERS (uniSERS) Coat the sample with metal, instead of adsorbing the sample to metal.
- Coat with metal (Magnetron sputtering). The sample bumbs out of plane of coating .
- Leave gap between metal cap and metal thin film
- The metal cap acts as an isolated oscillating electric dipole under electric field.
US12411069B2 - Method for determining the size of particles - Google Patents
3.4. Tip-Enhanced Raman Scattering
- TERS
- Variant of SERS
Synchronized with AFM (atomic force miscroscope)
SiN may be used as the tip. The concentrated field at the tip generates much higher signal.
3.5. Resonance Raman Scattering
- RRS
The excited state is chosen to be a real state. Amplifying the signal thousands to tens of thousands times.
4. Photoemission Spectroscopy
4.1. X-ray Photoelectron Spectroscopy
- XPS
The photoelectric effect at the topmost 5-10 nm of a surface is measured.
X-ray (1487.7 eV) from Al hits sample and the electron is separated by magnetic field. Electron flood gun can be pointed at the sample in order to neutralize it.
Peak position is related to chemical state, and peak area
5. Optical Detector
5.1. Filter
Undesired band are filtered out and passed onto the diffractor.
5.2. Grater
Grater is used to spead the wavelengths out.
5.3. CCD
CCD sensor is placed after a grater, which detect the position of each wavelength. Full vertical binning (FVB) may be used to sum across vertical pixels, and show the spectrogram.
6. X-Ray Diffraction
- XRD
- X-ray "pattern" is obtained
200 um - 1 mm of single crystal samples, or powder are used. Ring pattern is obtained from powder. While 2D diffraction pattern is observed for single crystal.
6.1. X-Ray Tube
Thermionic electron from cathode (usually tungsten) hits anode of metal target (usually copper, sometimes molybdenum). The anode then emits X-ray including its characteristic lines of anode and Bremsstrahlung. The X-ray is then emitted through beryllium window.
Anode needs cooling usually with water.
The characteristic lines are used in analysis. Usually, 154 pm \(\rm K_{\alpha}\) in copper, 71 pm \(\rm K_{\alpha}\) in molybdenum.
6.2. Bragg Equation
\[ n\lambda = 2d_{hkl}\sin\theta. \]
6.3. Powder Diffraction
Use powder. Powder is collection of microcrystals. Each crystal is like angle-dependent mirror. Think of radiance field. The crystal oriented in a way that the angle of incident matches the reflection condition for a crystal plane reflects x-ray. Giving overall pattern of concentric rings.
Calculate the interplanar spacings, or d-value, from the angle. Guess the Miller indices using reference or simulation. The lattice parameters can be obtained from them.
In cubic system, peaks with specific orders are observed:
| (hkl) | \(h^2 + k^2 + l^2\) |
|---|---|
| (100) | 1 |
| (110) | 2 |
| (111) | 3 |
| (200) | 4 |
| (210) | 5 |
| (211) | 6 |
| (220) | 8 |
| (300) = (221) | 9 |
X-ray pattern is used as fingerprint. There are public databases:
- ICDD. The International Centre for Diffraction Data | Materials Identification with the Powder Diffraction File -
- COD. Crystallography Open Database
Larger the crystal, higher SNR. Large single crystal can be obtained by solidified in ampule using Britmann method.
6.4. Debye-Scherrer Fomula
- Scherrer Equation
The relation between peak width and the size of crystal: \[ T = \frac{C\lambda}{B\cos\theta} = \frac{C\lambda}{\sqrt{B_M^2 - B_S^2}\cos\theta} \] where \(C \approx 0.9\text{ to }1.0\) is Scherrer constant, \(\lambda\) is the wavelength of x-ray, \(B\) is the FWHM(반칙폭) of the peak corrected for instrumental broadening: \(B_M\) is measured FWHM, \(B_S\) is the standard FWHM.
The formula is fine for 100 nm, but shaky for 100 nm to 200 nm, and become nonsense after 200 nm.
The diffraction pattern can be observed in time-series, as the crystal is being heated.
6.5. Rietveld Analysis
Simulating x-ray pattern?
7. X-Ray Spectroscopy
7.1. X-Ray Fluorescence Spectrometry
- X선 형광분석
Elements after Be emits x-ray with different wavelength when x-ray hits it. The x-ray emission is similar to the one caused by thermionic electron. Elements after Al is measured with confidence.
The x-ray fluorescence is used to measure the elemental composition and concentation, non-destructively. The concentration is measured semi-quantitatively; it can be measured but not reliable.
Largely divided into tow types: WD-XRF (wavelength dispersive) and ED-XRF (energy dispersive).
7.2. X-Ray Absorption Spectroscopy
- XAS
XANES (janes), EXAFS region
Measured at synchrotron.
XANES measure the absorption of the secondary x-ray. EXAFS measure the interference between nearby atoms.
The local environment of atoms can be measured this way.
8. Optical Microscopy
8.1. Bright-Field and Dark-Field
Bright-Field (BF) microscopy measure the trasmitted light directly, while Dark-Field (DF) microscopy blocks the direct light on the optical axis and measure the scattering of the light from the edge.
Figure 1: Semiconductor Defect Detection with Bright Field Microscopy and Dark Field Microscopy (deepblock.net)
8.2. DIC
위상차를 이용한 세포 구조 분석
9. Electron Microscopy
Scanning Electron Microscopy (SEM) is more common and cheap than Tunneling Electron Microscopy (TEM).
Billon dollars.
| SEM | TEM | |
| Resolution | ~10 nm | ~0.1 - 1 nm |
| Voltage | 1 - 30 kV | 100 - 300 kV |
| Detected Particle | Secondary Electron (SE) | Transmitted Electron |
| Back-Scattered electron (BSE) | Diffracted Electron |
Electron beam is generated from W filament, or LaB₆ filament.
TEM sample is placed on a grid of copper wire with diameter of 3 mm.
X-ray can be detected alongside to distinguish elements.
The maximum resolution can depend on the operator.
9.1. Scanning Electron Microscopy
- SEM
The amount of secondary electron determine the brightness in the image.
9.1.1. SEM-EDX
- Energy Dispersive X-ray Analysis
The x-ray is also measured to produce "color" image. SEM may or may not come with EDX.
9.2. Transmission Electron Microscopy
- TEM
TEM has specialized operator.
Chiller, Vacuum Pump, AVR are accompanied with the main machine.
Lattice fringe (the line pattern of crystal planes) can be observed.
Spherical abberation correction is applied for modern machines.
9.3. Cryo-TEM
2017 Nobel prize
The structure of protein is hard to make into a single crystal form.
Vitrify protein under cryogenic condition, and do 3D reconstruction of the molecule from the projection at various angles.
9.4. Focused Ion Beam
- combined with SEM and TEM
The sample is manipulated under electron microscope.
9.5. Ultra-High Voltage Electron Microscopy
1 MV - 3 MV
About 3 floors tall.
9.6. Scanning Transmission Electron Microscopy
- STEM
There exists multiple annular detector underneath. HADDF (high-angle annular dark-field imaging) images are obtained.
9.7. Electron Tomography
Image reconstruction method is applied to electron microscopy. Currently under development in KAIST.
10. Scanning Probe Microscopy
- SPM
The piezoceramic is used to control the microscopic motion.
10.1. Atomic Force Microscope
- AFM
Manufacturer: Park System
The cantilever is controled by piezoelectric scanner, and the displacement is read out using laser.
- Lateral resolution: 1 nm
- Depth resolution: 0.1 nm
10.1.1. Nano-IR Spectrometer
The surface is locally exited by IR and the tip detect photo-induced force (PiF)
10.2. Near-Field Scanning Optical Microscope
- NSOM
Metal-coated fiber optic is used to do spectroscopy at the surface itself.
10.3. Scanning Tunnelling Microscope
- STM
Voltage is applied between a metal wire tip and sample. The tunneling current is measured.
- Lateral resolution: 0.1 nm
- Depth resolution: 0.01 nm
- Require the sample to be conductive