Imaging

Table of Contents

1. X-Ray Imaging

1.1. X-Ray Microscope

1.2. X-Ray Computed Tomography

Sinogram is obtained by projecting in each angle \(\theta\). Projection at a single angle arise from the attenuation of x-ray radiated from a point source.

The raw data is preprocessed to account for the motion of the source, faulty detector, and noise.

1.2.1. Detector

Photon counting CT (PCCT) is currently tested for its sensitivity. The dead time problem need to be solved.

The traditional CT accumulates all the energy for a specified amount of time.

2. Magnetic Resonance Imaging

2.1. Components

Magnets Create a homogeneous magnetic field \(\vec{B}_0\) as strong as 0.5 T.

Gradient Coils Create gradient in the magnetic field, in the \(\vec{B}_0\) direction with a desired amount.

RF Coils

  • Magnetic field of the RF signal is called \(B_1\).
  • Send and receive RF signals in the traverse direction.
  • Not spacially selective.

2.2. Procedure

  • Utilize the Hydrogen nuclei.
  • Excite the nuclei with a radio wave of Larmor frequency
  • Measure the signal after echo time (TE) when the difference between tissues is at its maximum.
  • Re-exite the nuclei after repetition time (TR).

\[ S(\mathrm{TR}, \mathrm{TE})= (1-e^{-\mathrm{TR}/T_1})M_0e^{-\mathrm{TE}/T_2} \]

2.3. T₂ Weighted Image

\(\mathrm{TR}\gg T_1\) and \(\mathrm{TE} \sim T_2\)

2.4. T₁ Weighted Image

\(\mathrm{TR}\sim T_1\) and \(\mathrm{TE} \ll T_2\)

2.5. Spin Density Image

  • \(\mathrm{TR}\gg T_1\) and \(\mathrm{TE} \ll T_2\)
  • MRI measured with the echo time of zero which only reflects the proton density.

2.6. Hyperpolarization

  • Happen with ^{3}He, ^{13}C, ^{129}Ze, …
  • It can be used to increase the signal strength significantly.

2.7. Gradient Recalled Echo

2.7.1. Magnetic Field Gradient

\[ \frac{\partial B_z}{\partial x_i} = G_{x_i} \] Now \(\omega(\mathbf{x}) = \gamma (B_0+\mathbf{x}\cdot \nabla B_z)\)

Spins whose Larmor frequency matches the RF signals will experience excitation.

2.7.2. Slice Selection

Send a \(\rm sinc\) pulse that has transmit bandwidth \(\Delta \omega\), then the slice thickness is \[ \Delta z = \frac{\Delta \omega}{\gamma G_z} \]

A pulse looks like \[ B_1(t)=e^{i\omega_0 t}\mathrm{sinc}(\Delta \omega t) \] where \(\omega_0\) is the Larmor frequency at the center of the slice.

This is the .

The signal from the 2D slice \(R\) is \[ S(t)=\iint_R M(x,y)e^{i\phi(x,y,t)}\,dxdy \] where \(M(x,y)\) is the magnetization at TE, and \[ \phi(x,y,t)=-\gamma\left[\int_0^t G_x(u)x\,du + \int_0^{t} G_y(u) y\, du\right]. \]

2.7.3. Phase Encoding

Impose gradient \(G_y\) for time \(\tau\), and have \[ \phi(x,y,t)=-\gamma\left[\int_0^t G_x(u)x\,du + \tau G_y y\right] \]

2.7.4. Frequency Encoding

Impose gradient \(G_x\) while doing the read out, and get \[ S(t)=\iint_R M(x,y)e^{-i\gamma (tG_x x+\tau G_yy)}\, dxdy \] where \(t\) starts at the start of \(G_x\).

This is a line of known phasor in \(k\)-space.

3. Image Reconstruction

3.1. Sinogram

Sinogram is a sequence of one-dimensional projection around a fix point. Each point in space would appear as a sine curve in sinogram.

Sinogram is described by Radon transform of density field \(f(x,y)\):

\begin{align*} p(s, \phi) &= \mathcal{R}{f(x,y)} \\ &= \int_{L_{s,\phi}} f(x,y)\dd{u} \\ &= \int_{-\infty}^{\infty} f(s\cos\phi - u\sin\phi, s\sin\phi + u\cos\phi)\dd{u}. \end{align*}

3.2. Simple Back Projection

The attenuation of a line is given by \[ N = N_0e^{-\mu x}. \]

Each constant \(\theta\) line in the sinogram is spread across in its angle, and added up.

3.3. Filtered Back Projection

Does the .