Nuclear Medicine

Table of Contents

Broadly single photon imaging (mainly single photon emission computed tomography SPECT) and positron imaging (positron emission tomography PET). Magnetic resonance imaging (MRI) and x-ray computed tomography (CT) are not part of it.

Planar image (projection) and tomographic reconstruction from images from multiple angles.

Advantages

1. Pulse-Height Spectrometry

Some detectors provide pulse-height proportional to the energy of the radiation energy.

  • Proportional Counter
  • Scintillation Detetor
  • Semiconductor Detector

The pulse-height spectrum depends on

  • the energy spetrum of radiation source
  • type of radiation detector
  • mechanisms by which the radiation energy is deposited

The energy of a single particle is deposited by ionizing primary particles which in turn ionize secondary particles. Some particle can escape without fully depositing the energy, when the detector is not thick enough. Photons are especially vulnerable at losing some energy.

2. Tracer Principle

tracer principle (described in 1913 and tested in 1923 by George de Hevesy) states that a minute, detectable amount of a labeled substance (a radioisotope or stable isotope) can be introduced into a system to follow the path, distribution, or metabolic behavior of its normal chemical counterpart without disturbing or altering the system itself.

3. Binding Potential

3.1. Quantities

Free concentration \(C_{\rm F}\) is the concentration of radiotracers that are not bound to any molecule.

Bound concentration is divided into two: specifically bound \(C_{\rm S}\) and nonspecifically bound \(C_{\rm NS}\). Specifically bound tracer is the one bound to the target of interest.

Nondisplaceable concentration is the sum of free and nonspecifically bound concentration: \[ C_{\rm ND} = C_{\rm F} + C_{\rm NS}. \]

Free fraction \(f_{\rm ND}\) of nondisplacement compartment means \[ f_{\rm ND} = \frac{C_{\rm F}}{C_{\rm ND}}. \]

In tracer-limit approximation, the receptor-ligand equilibrium would be \[ C_{\rm S} = \frac{B_{\rm avail}}{K_D} C_{\rm F}. \] since the dissociation equilibrium constant between target receptor and radioligand would be approximated by: \[ K_D = \frac{C_{\rm F} B_{\rm avail}}{C_{\rm S}} \]

3.2. Binding Potential Relative to Nondispalceable Uptake

  • \(\rm BP_{ND}\)

At equilibrium, \[ \mathrm{BP}_\mathrm{ND} = \frac{C_{\rm S}}{C_{\rm ND}} = \frac{f_{\rm ND} B_{\rm avail}}{K_D} \] where \(C_\mathrm{S}\) is the concentration of specifically bound radiotracer in tissue, \(C_{\rm ND}\) is the concentration of nondispalceable radiotracer, including free and nonspecifically bound tracer, \(f_{\rm ND}\) is the free fraction of radiotracer in the nondispalceable tissue compartment, \(B_{\rm avail}\) is the concentration of available binding sites, and \(K_D\) is the equilibrium dissociation constant of the radiotracer-target interaction.

At equilibrium, the total distribution volume is \[ V_{\rm T} = \frac{C_{\rm T}}{C_{\rm P}}, \] and nondispalceable distribution volume is \[ V_{\rm ND} = \frac{C_{\rm ND}}{C_{\rm F}}. \] The binding potential can be rewritten as \[ \mathrm{BP}_{\rm ND} = \frac{V_{\rm T}}{V_{\rm ND}} - 1 = \mathrm{DVR} - 1. \] where \(\rm DVR\) is the distribution volume ratio.

3.3. Receptor Occupancy

Occupancy \(\theta\) of a specific receptor is \[ \theta = \frac{B_{\rm bound}}{B_{\rm total}} \] where \(B_{\rm bound}\) is the concentration of bound receptor, and \[ B_{\rm total} is the total concentration of receptor. \]

When a single kind of ligand \(L\) binds independently to the binding sites of receptor R + L ↔ RL \[ \theta = \frac{[L]}{K_D + [L]}. \]

Occupancy can be expressed with binding potential before and after administering a competing drug \[ \theta = 1 - \frac{\mathrm{BP}_{\rm ND}^{\rm drug}}{\mathrm{BP}_{\rm ND}^{baseline}}. \]

High occupancy is pharmacologically, and physiologically significant, but it does not necessarily means high radiation exposure.